Open Access Journal Article

Performance of breath-actuated (Qvar) pressurize metered dose inhaler with different spacers: In-vitro study

by Mohammed Almeziny 1,* Brian Clark 2  and  Henry Chrystyn 3
1
Consultant Clinical Pharmacist, Prince Sultan Military Medical City (PSMMC), Riyadh, KSA
2
Institute of Pharmaceutical Innovation, School of Pharmacy, University of Bradford, UK
3
Inhalation Consultancy Ltd, Leeds, UK
*
Author to whom correspondence should be addressed.
IJCMR  2025 3(3):59; https://doi.org/10.61466/ijcmr3030004
Received: 3 March 2025 / Accepted: 29 April 2025 / Published Online: 4 May 2025

Abstract

Background

Spacers are commonly used especially for paediatric and elderly patients. However, there is a considerable discussion about their use and operation. The purpose of this study was to examine the performance of different types of spacers with Qvar breath actuated pressurize metered dose inhaler (pMDI) containing Beclometasone with fine particles.

Methods

The dose evaluation method for pMDIs in the British Pharmacopoeia (BP) was used to find out the total emitted dose (TED) and particle size distribution of Qvar pMDI. The study looked at pMDI alone and with three different spacers (Optimizer, Aerochamber Plus, and Aerochamber MAX) at a flow of 28.3 L/min and 4 L inhalation volume to mimic adults. At different times, from the first to the last actuation, the average amount of drug in each pMDI alone and connected to the three different spacers was evaluated.

Results

The pMDI by itself had the highest TED, which was significantly higher than all pMDI-spacer combinations (p < 0.05). The spacers also had higher fine particle dose, and fine particle fraction (p<5 µg) then pMDI alone. No significant different was observed in mass median aerodynamic diameter between the four tested combinations. No significant difference was observed between the three evaluated spacer.

Conclusions

These results show that there are improving effects on aerodynamic characterisation and the amount of drug available for inhalation when spacers are used as inhalational aids to pMDI. That support the GINA recommendation regarding the use of spacer with all patients using pMDI even the breath actuated pMDIs.


Copyright: © 2025 by Almeziny, Clark and Chrystyn. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY) (Creative Commons Attribution 4.0 International License). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

Share and Cite

ACS Style
Almeziny, M.; Clark, B.; Chrystyn, H. Performance of breath-actuated (Qvar) pressurize metered dose inhaler with different spacers: In-vitro study. International Journal of Clinical Medical Research, 2025, 3, 59. doi:10.61466/ijcmr3030004
AMA Style
Almeziny M, Clark B, Chrystyn H. Performance of breath-actuated (Qvar) pressurize metered dose inhaler with different spacers: In-vitro study. International Journal of Clinical Medical Research; 2025, 3(3):59. doi:10.61466/ijcmr3030004
Chicago/Turabian Style
Almeziny, Mohammed; Clark, Brian; Chrystyn, Henry 2025. "Performance of breath-actuated (Qvar) pressurize metered dose inhaler with different spacers: In-vitro study" International Journal of Clinical Medical Research 3, no.3:59. doi:10.61466/ijcmr3030004

Article Metrics

Article Access Statistics

References

  1. Saeed, H., Abdelrahim, M.E., Rabea, H., et al., Evaluation of Disposable and Traditional Accessory Devices for Use With a Pressurized Metered-Dose Inhaler. Respir Care, 2020. 65(3): p. 320-325.
  2. Saeed, H., Abdelrahim, M.E., Rabea, H., et al., Evaluation of disposable and traditional accessory devices for use with a pressurized metered-dose inhaler. Respiratory care, 2020. 65(3): p. 320-325.
  3. Nelson, H.S., Inhalation devices, delivery systems, and patient technique. Annals of Allergy, Asthma & Immunology, 2016. 117(6): p. 606-612.
  4. Cazzola, M., Rogliani, P., Anzueto, A., et al., Inhaled medication: which device for which patient. ERS Monogr, 2015. 69: p. 213-223.
  5. Steiropoulos, P., Bakakos, P., Hatziagorou, E., et al., The present and future of inhalation therapy for the management of obstructive airway diseases: Emphasis on pressurized metered-dose inhalers. Pneumon, 2021. 34(4): p. 24.
  6. Moraga-Espinoza, D.F., Brunaugh, A.D., Ferrati, S., et al., Overview of the delivery technologies for inhalation aerosols, in Inhalation Aerosols. 2019, CRC Press. p. 123-143.
  7. Bisgaard, H., Smaldone, G.C., and O'Callaghan, C., Drug Delivery to the Lung. Vol. 162. 2002, New York Marcel Dekker. 511.
  8. Asthma., G.I.f. Global Strategy for Asthma Management and Prevention 2024. 2024 October 31 2024.]; Available from: https://ginasthma.org.
  9. Saeed, H., Ammari, W.G., Osama, H., et al., Evaluation of a cardboard-based spacer for enhancing aerosol delivery from pressurized metered dose inhalers. Journal of Asthma, 2025(just-accepted): p. 1-9.
  10. Rahmatalla, M.F., Zuberbuhler, P.C., Lange, C.F., et al., In Vitro Effect of a Holding Chamber on the Mouth-Throat Deposition of QVAR®(Hydrofluoroalkane-Beclomethasone Dipropionate). Journal of Aerosol Medicine, 2002. 15(4): p. 379-385.
  11. Anderson, G., Johnson, N., Mulgirigama, A., et al., Use of spacers for patients treated with pressurized metered dose inhalers: focus on the VENTOLIN™ Mini Spacer. Expert Opinion on Drug Delivery, 2018. 15(4): p. 419-430.
  12. Nikander, K., Nicholls, C., Denyer, J., et al., The evolution of spacers and valved holding chambers. Journal of aerosol medicine and pulmonary drug delivery, 2014. 27(S1): p. S-4-S-23.
  13. Lung, N.H. and Institute, B., Expert panel report 3 (EPR3): guidelines for the diagnosis and management of asthma. http://www. nhlbi. nih. gov/guidelines/asthma/, 2007.
  14. Nicola, M., Hussein, R.R., Soliman, Y.M., et al., Effect of different accessory devices on the dose delivered from pressurised metred‐dose inhalers. International Journal of Clinical Practice, 2021. 75(6): p. e14157.
  15. Nicola, M., Soliman, Y.M., Hussein, R., et al., Comparison between traditional and nontraditional add-on devices used with pressurised metered-dose inhalers. ERJ Open Research, 2020. 6(4).
  16. Saini, D., Biris, A., Srirama, P., et al., Particle size and charge distribution analysis of pharmaceutical aerosols generated by inhalers. Pharmaceutical development and technology, 2007. 12(1): p. 35-41.
  17. Ganderton, D., Lewis, D., Davies, R., et al., Modulite®: a means of designing the aerosols generated by pressurized metered dose inhalers. Respiratory Medicine, 2002. 96(Supplement 4): p. S3-S8.
  18. Roland, N.J., Bhalla, R.K., and Earis, J., The Local Side Effects of Inhaled Corticosteroids* Current Understanding and Review of the Literature. Chest, 2004. 126(1): p. 213-219.
  19. Product-information, Qavr® product monograph 2006, 3M Drug Delivery Systems.
  20. Burgt, J.A.V., Busse, W.W., Martin, R.J., et al., Efficacy and safety overview of a new inhaled corticosteroid, QVAR (hydrofluoroalkane-beclomethasone extrafine inhalation aerosol), in asthma. Journal of Allergy and Clinical Immunology, 2000. 106(6): p. 1209-1226.
  21. Hardy, J.G., Jasuja, A.K., Frier, M., et al., A small volume spacer for use with a breath-operated pressurised metered dose inhaler. International Journal of Pharmaceutics, 1996. 142(1): p. 129-133.
  22. Product-Monograph, Aerochamber Max Product monograph, 2006, Trudell Medical International (TMI).
  23. Product-Monograph, Aerochamber Plus Product Monograph, 2005, Trudell Medical International (TMI).
  24. Almeziny, M., Performance of two different types of inhalers. Influence of flow and spacer on emitted dose and aerodynamic characterisation., 2009, University of Bradford.
  25. Barry, P.W. and O'Callaghan, C., Inhalational drug delivery from seven different spacer devices. Thorax, 1996. 51(8): p. 835.
  26. Nagel, M.W., Wiersema, K.J., Bates, S.L., et al., Performance of Large-and Small-Volume Valved Holding Chambers with a New Combination Long-Term Bronchodilator/Anti-inflammatory Formulation Delivered by Pressurized Metered Dose Inhaler. Journal of Aerosol Medicine, 2002. 15(4): p. 427-433.
  27. Abdelfattah, A.M., Sarhan, R.M., Madney, Y.M., et al., User-Friendliness Evaluation of Handling pMDI with Various Add-on Devices in Asthmatic Patients. AAPS PharmSciTech, 2024. 25(8): p. 274.
  28. Devadason, S.G., Recent Advances in Aerosol Therapy for Children with Asthma. Journal of Aerosol Medicine, 2006. 19(1): p. 61-66.
  29. Saeed, H., G., A.W., Hasnaa, O., et al., Evaluation of a cardboard-based spacer for enhancing aerosol delivery from pressurized metered dose inhalers. Journal of Asthma, 2025: p. (in press).
  30. Terzano, C., Pressurized Metered Dose Inhalers and Add-on Devices. Pulmonary Pharmacology & Therapeutics, 2001. 14(5): p. 351-366.
  31. Anhoj, J., Bisgaard, H., and Lipworth, B.J., Effect of electrostatic charge in plastic spacers on the lung delivery of HFA-salbutamol in children. British Journal of Clinical Pharmacology, 1999. 47(3): p. 333.
  32. Geller, D. and Kesser. Effects of Pre-treatment of Static and Non-static Holding Chambers (HC) on the Delivery of Flovent 110-HFA (F) in an Infant Airway Model. in The American Thoracic Society Conference. 2006. San Diego, USA.
  33. Iula, A.K., Obetz, C.L., and Deluccia, F., Comparative study of dose delivery from an Azmacort™ MDI paired with an Azmacort™ integrated actuator/spacer, an Aerochamber™, Inspirease™, and a Microspacer™. Respir Drug Deliv, 1996: p. 352–353.
  34. Nicola, M., Hussein, R.R.S., Soliman, Y.M.A., et al., Effect of different accessory devices on the dose delivered from pressurised metred-dose inhalers. International Journal of Clinical Practice, 2021. 75(6): p. e14157.
  35. Asmus, M.J., Liang, J., Coowanitwong, I., et al., In vitro deposition of fluticasone aerosol from a metered-dose inhaler with and without two common valved holding chambers. Annals of Allergy, Asthma and Immunology, 2002. 88(2): p. 204-208.
  36. Cripps, A., Riebe, M., Schulze, M., et al., Pharmaceutical transition to non-CFC pressurized metered dose inhalers. Respiratory Medicine, 2000. 94(Supplement 2): p. S3-S9.
  37. Rabe, K.F., Hurd, S., Anzueto, A., et al., Global Strategy for the Diagnosis, Management, and Prevention of Chronic Obstructive Pulmonary Disease: GOLD Executive Summary. American Journal of Respiratory and Critical Care Medicine, 2007. 176(6): p. 532.
  38. Nicola, M., Soliman, Y.M.A., Hussein, R., et al., Comparison between traditional and nontraditional add-on devices used with pressurised metered-dose inhalers. ERJ Open Research, 2020. 6(4): p. 00073-2020.
  39. Leach, C.L., Davidson, P.J., Bredow, T.S., et al., Patient factors influencing the deposition of HFA-beclomethasone dipropionate metered dose inhaler. Eur Respir J, 1998. 12: p. 66s.
  40. Fink, J.B., Wilkes, W., Dhand, R., et al., Do spacers and holding chambers influence respirable mass from a metered-dose inhaler. Am J Respir Crit Care Med, 1998. 157(3).
  41. Asmus, M.J., Liang, J., Coowanitwong, I., et al., In Vitro Performance Characteristics of Valved Holding Chamber and Spacer Devices with a Fluticasone Metered-Dose Inhaler. Pharmacotherapy, 2004. 24(2): p. 159-166.
  42. Faarc, D., Csci, J., and Nagel, M.W., Levalbuterol Aerosol Delivery With a Nonelectrostatic Versus a Nonconducting Valved Holding Chamber. Respiratory Care, 2006. 51(5): p. 511.
  43. Rau, J.L., Coppolo, D.P., Nagel, M.W., et al., The importance of nonelectrostatic materials in holding chambers for delivery of hydrofluoroalkane albuterol. Respiratory care, 2006. 51(5): p. 503-510.
  44. Tony, S.M., A., A.M., Mogeda, A.E., et al., Effect of using acoustic flo-tone training device and its smartphone application on enhancing inhalation technique from metered-dose inhaler with spacer in asthmatic children. Experimental Lung Research, 2022. 48(7-8): p. 224-238.